Method 81

Crotonaldehyde by HPLC


This method describes the sample collection and analysis of crotonaldehyde. A sample is collected by drawing air through an open face air monitoring cassette containing two glass fiber filters, each of which is coated with 2,4-dinitrophenylhydrazine and phosphoric acid. The sample is extracted with acetonitrile and analyzed by HPLC using a UV detector. Store samples at -20°C upon receipt at the laboratory. If such storage is not possible, samples must be analyzed within 9 days after collection. Keep the samples in the dark whenever possible as a precaution against photodecomposition.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   At least 18 days when stored at -20°C. Samples must be analyzed within 9 days after collection if stored at room temperature.
Preservatives:   Store samples at -20°C upon receipt at the laboratory. Keep the samples in the dark whenever possible as a precaution against photodecomposition.
Required Preps:   Two DNPH and and phosphoric acid coated glass fiber filters contained in an open face air monitoring cassette.
Collection Method:   Treated glass fiber filter: 6 L at 0.1 L/min.
Analytical Methodology:   HPLC/UV.
Documentation:   081

Analyte List*

Analyte Formula CAS Number Detection Limit
Crotonaldehyde
C4H6O
123-73-9
0.56
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.