Method PV2048

Phenothiazine


This method is for the determination of phenothiazine. The sample is collected on a 37mm glass fiber filter. After collection, the sample is extracted with methy tert-butyl ether and analyzed by GC/NPD (gas chromatography/nitrogen phosphorus detector). After collection and during shipment the samples should be protected from light.


(OSHA)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   The method notes that the recoveries of the analyte were above 96% after storage at ambient and refrigerated temperatures for 14 days.
Preservatives:   Protect samples from light.
Required Preps:   37 mm glass fiber filter, SKC 225-706
Collection Method:   The method recommends a flow rate of 1 L/min. with a 100L collection volume.
Analytical Methodology:   GC/NPD
Documentation:   PV2048

Analyte List*

Analyte Formula CAS Number Detection Limit
Phenothiazine
C12H9NS
92-84-2
2
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.