Allyl Glycidyl Ether by GC/FID


This method is for the analysis of allyl glycidal ether. The sample is collected on Tenax® sorbent tubes, desorbed in diethyl ether, and analyzed by GC/FID. The working range is 2 to 20 ppm (10 to 100 mg/m³) for a 3L air sample. An appropriate capillary column may be used for better resolution and sensitivity. NOTE: This is method S346 in a revised format.


(NIOSH Issue 1: 15 August 1994)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   At least 7 days @ 25 °C
Preservatives:   25 °C
Required Preps:   Solid sorbent tube: Tenax® GC 100mg/50mg; 10cm x 8mm OD x 6mm ID
Collection Method:   Solid sorbent tube (SKC 226-35-03) 1.5 L to 8 L @ 0.01 to 0.2 L/min
Analytical Methodology:   GC/FID
Documentation:   2545

Analyte List*

Analyte Formula CAS Number Detection Limit
Allyl Glycidyl Ether
C6H10O2
106-92-3
10
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.