Chlorinated camphene (Toxaphene) by GC/ECD


This method is for the determination of chlorinated camphene (toxaphene). The sample is collected on cellulose ester membrane filters, desorped in 10mL petroleum ether, and analyzed by GC with a 63Ni electron capture detection system. The working range is 0.05 to 1.5 mg/m³ for a 15-L air sample. Note: This is Method S67 in a revised format.


(NIOSH Issue 1: 15 August 1994)

Request-A-Quote

No Obligation Quotation for Analytical Services

If you would like us to provide a quote for laboratory analysis, just provide us with as much information as you can about your project (the more, the better) and we'll provide you a quote via email. As you are searching or browsing our Analytical Guide, you'll see the Request-A-Quote icon... just click on it to start the request process.
Request-A-Quote

Method Data

Hold Times, Preservatives, Preps, Collection, Analytical & Documentation
Holding Time:   Not determined.
Preservatives:   None noted in method.
Required Preps:   Cellulose ester membrane filter; 0.8-µm.
Collection Method:   Cellulose ester membrane filter: 2L - 30L @ 0.2 - 1L/min.
Analytical Methodology:   GC/63Ni ECD
Documentation:   5039

Analyte List*

Analyte Formula CAS Number Detection Limit
Toxaphene
C10H10Cl8
8001-35-2
0.14
 µg

* The analytes and detection limits listed for each method represent the typical detection limits and analytes reported for that particular method. Keep in mind that analyte lists may vary from laboratory to laboratory. Detection limits may also vary from lab to lab and are dependent upon the sample size, matrix, and any interferences that may be present in the sample.